| Item |
|
| CV-100AT CV100 |
CV700AT CV700 |
| Load |
Type A |
| Kgf |
1 |
2 |
3 |
5 |
10 |
20 |
30 |
50 |
| N |
9.807 |
19.61 |
29.42 |
49.03 |
98.07 |
196.1 |
294.2 |
490.3 |
|
| Type B |
| Kgf |
0.3 |
0.5 |
1 |
3 |
5 |
10 |
20 |
30 |
| N |
2.942 |
4.903 |
9.807 |
29.42 |
49.03 |
98.07 |
196.1 |
249.2 |
|
| Loading Mechanism |
Automatic loading and releasing method |
| Load Applying speed |
Initial loading speed: 120µ m/sec---Actual loading speed: 6-µ m/sec (automatic speed changing) |
| Dwell time |
5~40 sec |
5~99 sec |
| Turret Mechanism |
Automatic /Manual |
Automatic/Manual |
| Object Lens Mounting |
Type A: x10 Type B: x10 & x20 (optional: max. 3 lenses) |
Magnification of Microscope
|
Type A: x100 (Objective 10 x Eyepiece 10) Type B: x100 (Objective 10 x Eyepiece 10) & x200 (Objective 20 x Eyepiece10)
|
| Measuring Microscope |
Mechanical |
Electronic |
| Max. Measuring Length |
x 100: 800µm x 200: 400µm |
| Min. Graduation |
Micrometer : 0.5µm |
Digital : 0.1µm |
| Min. Measuring Unit |
Eye Judgment : 0.1µm |
Digital : 0.1µm |
| LCD Touch Panel Display
[Operation & Data Display]
|
|
Operation & Selection Test Conditions
Specimen No. (numerals & alphabet), Print out modes.
Magnification, RS232C interface, dwell time, bright of light source, loading start, etc.
Data Display (Explanation of functions by icon) Diagonal length (D1/D2), Hardness value (HV, HK, HBS, & HBW) & fracture toughness (Kc), Conversion value, Test load (N or Kgf), Test times, OK/NG criteria, Data editing, (Data memory): Max. 999 test datas), Self diagnosis (Display troubles with motor & switches), etc.
|
| Max. Height of Specimen |
210mm |
| Max. Depth of Specimen |
165mm |
| Available Test Parameter |
|
HV, HK, HBS, HBW, & Kc (Fracture toughness Value) |
| Brinell Hardness |
|
Calculation for HBS & HBW (Optional small steel ball indenters and special test loads)
|
| Fracture Toughness Value(KC) |
|
Conforms to JIS R 1507/f1 method |
| Hardness Conversion |
|
Conforms to SAE (J-417b) & ASTM (E-140) |
| Data Memory |
|
Max. 999 test datas |
| Data Output |
|
Rs232C: Baud Rate 300-~19200bps (factory setting: 9600bps)
CENTRONICS parallel or Serial output selectable (Factory Setting: CENTRONICS)
I/O fir external control : Neg. logic O/C (open collector), 5~12V.
|
| OK/NG Criteria |
|
Upper and lower limits setting and HI/OK/LO |
| Self Diagnosis |
|
Displays troubles with motors and switches. |
| Light Source |
6V 18W |
6V 18W: Automatic brightness reduction during loading |
| Optical Function |
Field Aperture : Changeable
Numerical Aperture : Changeable
Color filter : Changeable
|
| Data Editing {Capable to edit after completion of measurement program)
|
|
Measurement data editing (correction, deleting)
Conversion data display
Batch conversion of memorized measurement data
Statistic data display:
Mean value
Max. Value
Min. value
Standard Deviation
Printout after data editing:
Measurement Data
Conversion Value, max. value, min. value, dispersion (R)mean value, standard deviation
|
| RS232C Output & Printout Mode |
|
All data (HV, HK, test load, diagonal length, conversion scale, evaluation results)
Measurement
|
| Accuracy |
Conform to JIS B-7725 and ISO 146 |
| Dimensions |
Approx. W 195 x D 517 x J 690 mm |
| Weight |
Approx. 55 kg (including standard accessories) |
| Power Supplies |
Single phase AC 100~240V 50 / 60HZ |